TUNNEL-TYPE ASEPTIC TRANSFER CHAMBER

EN EN

TUNNEL-TYPE ASEPTIC TRANSFER CHAMBER

The system is primarily used in laboratory animal facilities for transferring and biodecontaminating animal cages, materials and supplies. Hydrogen peroxide or ozone is used for disinfection.

Detail

FEATURES

• SUS316L construction, 3 mm thick; 8 mm tempered-glass viewing window.

• ISO 4.8 cleanliness; comprehensive 4Q qualification services.

• Integrated temperature, humidity, differential pressure and H₂O₂ concentration monitoring.

• Online pressure-decay leak testing, meeting ISO 10648, ISO 14644 and PDA TR34.

• Design complies with GMP, ISO, ISPE and pharmacopoeial requirements.

• PLC and touchscreen provide a user-friendly human–machine interface.

• Password protection or three-level access control; only authorized personnel may run biodecontamination or set parameters. Embedded software supports upgrades and maintenance.

• Recipe management: select biodecontamination programs for different material types (optional).

• Scheduled automatic startup.

• Stores sterilization and process parameters; exports Excel-format data through USB. An integrated dot-matrix printer prints biodecontamination data in real time.

• Equipment, temperature and pressure alarms ensure process reliability.

• Integrated UV lamp with irradiance ≥ 100 µW/cm².