CLEAN BENCH
A clean bench is designed to provide ISO 5 (Class 100) cleanliness in a local working area. Airflow may be horizontal or vertical unidirectional flow.
STANDARD CONFIGURATION
• LCD display and control
• EBM fan
• HEPA filter
• LED lighting
• Stainless steel worktop
• Swivel castors
OPTIONS
• Air velocity sensor
• Automatic PLC control
• Airborne particle and viable sampling
• Differential pressure gauge
TECHNICAL SPECIFICATIONS
• Cleanliness: Grade A (GMP)
• H14: 99.995%; two-stage prefiltration and HEPA filtration
• Air velocity: 0.36–0.54 m/s
• Leak rate: 0.01%
• Noise: ≤ 65 dB
• Illuminance: ≥ 300 lux
• SUS304 worktop
• Power supply: 220 V
APPLICATIONS
Electronics, optoelectronics, precision instruments, instrumentation, new energy, photovoltaics, pharmaceuticals, healthcare, microbiology, agricultural breeding research, food processing and laboratories.